This day i will taking about Furnace Camera.As the current
technological developments in power plant technology especially the use of
recovery boiler is constantly innovating in order to complement the existing
deficiencies in the means of production to facilitate the operation of the
recovery boiler.
Currently the power plant that uses a recovery boiler there
is a lack of technology to the unavailability of a device to control the
conditions inside the boiler (furnace). So far the recovery boiler
operator can only monitor the condition of the inside of the furnace manually
just by looking at the field by controlling the air of room pannel (dcs). but
that's not enough data taken with the actual
data. accuracy conditions may only be 65% to monitor conditions
inside the furnace charbed.
Camera Enclosure with Lens
The above shows what is removed and store during shutdown or
repair
Camera Enclosure Interior
Back of camera core
Removing Camera Core
Camera Retract system
Cleaner
Control Enclosure
Solenoid valve assembly
(Inside control enclosure)
Valve Manual override
Siemens LOGO PLC
Settable Parameter available from key pad on Logo:
Cycle time (time between cleaning cycles)
Cleaning stroke time (time energized and de-energized
Number of cleaning strokes
|
|
|
Pneumatic System
Lens tube/manifold pressure to be maintained between 1.5 and
2 BAR
Input pressure to the system should be 5 and 12 bar
The cleaning cylinder (with check valve) acts as an
accumulator to retract camera in the event of a line failure.
System Faults and General Maintenance
LENS CLEANING
The most common form of maintenance on the camera lens
system will be cleaning the “objective” lens – that part of the lens system
furthest from the camera. Periodic, daily cleaning of the objective lens should
be expected, although cleaning intervals of 5 or 6 days are not un-common.
A dirty objective will produce an image that is fuzzy or
appears out of focus. The part of the objective that requires cleaning is the
protective clear sapphire window, which is very hard and difficult to
inadvertently scratch, but relatively easy to break.
Warning:
Do not attempt to clean a hot lens! The objective lens
assembly must be below 110°F or comfortable to the touch before cleaning.
To clean the objective lens sapphire:
Retract the lens by selecting “RETRACT” at the control
enclosure.
Allow the lens to cool for several minutes.
Once cool (relatively comfortable to the touch), turn the
supply air to the lens system off at the shut-off valve.
Using a cotton swab and alcohol, reach into the end of the
lens assembly and clean any dirt or oil that may have collected on the
objective sapphire.
Char build-up on or around the lens tube should also be
cleaned at this time.
Turn the supply air on.
Most Faults and the probable reason for the fault are
broadcast on the face of the Control Panel
The camera has 30-seconds to cool down once an overtemp
Situation is discovered.
There are three different faults that will cause the camera
to retract:
Camera enclosure over temp - temporary over temp
recognized
Camera has retracted and then cooled to operating temp check
air supply and for combustion air leaks
Camera over temp - system is shut down
Camera retracted and failed to cool within 30 seconds camera
has been shut off as result to avoid damage. Check air supply and cooler
adjustment.
Low air pressure – camera retracted to avoid overheating
Check for compressed air leaks and ball valve position
Retract limit fail – automatic cleaning not possible
Important to keep the retract cleaned and lubricated so
if a overtemp condition occurs the retract can do what it is supposed to do
Output fail check fuse 1
Output fail check fuse 2
There are three different ouput faults that are monitored
Maybe usefull.
0 komentar:
Posting Komentar